Performs in-situ bio-decontamination of BIBO HEPA filters through a four-step cycle (dehumidify – condition – treat – aerate), developed to a ≥6-log sporicidal-reduction target against Geobacillus stearothermophilus.
Request a Quote View All Products
Overview
This VHP system is used for in-situ bio-decontamination of BIBO (Bag-In/Bag-Out) HEPA filters. After normal operation of the BIBO filter, the manual valve is opened and the VHP unit is started to run the cycle. The cycle has four steps — dehumidify, condition, treat and aerate: Dehumidify — the unit's built-in dehumidification function uses an internal desiccant (molecular sieve) to absorb moisture; Condition / Treat — VHP vapor is introduced to treat the BIBO HEPA filter, typically at a concentration of around 30 ppm for roughly 40 minutes, with the cycle developed to a ≥6-log sporicidal-reduction target against Geobacillus stearothermophilus, subject to validation; Aerate — residual VHP is broken down via the unit's internal palladium-catalyst recirculation loop, or by running the BIBO housing's own supply/exhaust system, bringing the hydrogen peroxide concentration down below 1 ppm.
Technical Specifications
| External dimensions (W×D×H) | 640×836×1,036 mm | Metal / non-metal materials | 304, 316L stainless steel / PVC, silicone |
|---|---|---|---|
| Power / frequency / rating | 220 V AC / 50 Hz / ≥3 kW | Touchscreen / control system | Siemens HMI / Haide tablet PC · Siemens PLC |
| Humidity / temperature / pressure resolution | 0.1% / 0.1°C / 0.1 Pa | Noise | ≤75 dB(A) |
| Air flow | ≤500 L/min | Treatment volume | ≤30 m³ |
| Vaporization temperature | ≤120°C | Dehumidification target | <30% RH |
| Sterilant | 30% or 35% hydrogen peroxide solution | Sporicidal efficacy | ≥6-log reduction (subject to validation) |
| Net weight | Approx. 80 kg |